A dynamometer-type wattmeter is used to measure the power consumed by…
Previous year question for Non-Conventional Energy Resources and Instrumentation.
Question
A dynamometer-type wattmeter is used to measure the power consumed by a single-phase inductive load.
The readings are as follows:
Voltage across the load = 230 V
Current through the load = 10 A
Power factor of the load = 0.75 lagging
Wattmeter has a 2.5% low reading error due to stray magnetic fields and friction
Wattmeter constant = 600 revolutions per kWh
Calculate:
a) The actual power consumed by the load.
b) The ideal wattmeter reading.
c) The wattmeter reading with the 2.5% error applied.
If the load runs for 5 hours, calculate the total energy consumed and the number of revolutions of the energy meter.
- Year: 2022
- Subject: Non-Conventional Energy Resources and Instrumentation